Humidifiers and Heated Wire Breathing Systems: Classification and Grouping in Costa Rica

respiratory humidifier in hospital, classification humidifiers Costa Rica

Share This Post

This article closes our series on medical devices related to respiratory therapy, where we have covered topics like anesthesia circuits, oxygen masks, and breathing systems. Now, we’ll explore the regulatory process for registering humidifiers and heated wire breathing systems (HWBS) in Costa Rica. 

 

What are Humidifiers and Heated Wire Breathing Systems?

 

  • Humidifiers: medical devices designed to add moisture and, often, heat to the inspired gases delivered to patients undergoing respiratory support to mitigate risk of complications, particularly during prolonged ventilation. 
  • Heated Wire Breathing Systems (HWBS): specialized type of breathing circuits that incorporate heated wires within the tubing to maintain the temperature of the delivered gases, preventing condensation within the circuit. 

 

These devices are used in conjunction with mechanical ventilators and various airway management devices (endotracheal tubes, laryngeal masks, face masks, etc.) in settings like surgery, emergency care, intensive care units (ICUs), and home healthcare. 

 

 

What are the components of a Humidifier and its interaction with a HWBS? 

 

Humidifiers and HWBS, while separate devices, work synergistically: 

  • Humidifier Components: 
    • Heating Element: Warms the water to create water vapor. 
    • Water Chamber: A reservoir holding sterile water. This can be a pre-filled, disposable chamber or a refillable chamber. 
    • Temperature Control System: Allows clinicians to adjust the temperature and, indirectly, the humidity level. 
    • Connection Ports: To connect to the breathing circuit (often the inspiratory limb). 

 

  • HWBS Components: 
    • Breathing Tubes: Flexible tubes connecting the patient’s airway to the oxygen source or ventilator. 
    • Heated Wires: Run along the inspiratory limb (and often the expiratory limb) of the breathing circuit. 
    • Temperature Sensors: Located within the circuit (often near the patient connection) to monitor gas temperature. 
    • Controller/Adapter: connects the heated wires to a power source from the ventilator and transmits sensor data. 

 

What types of Humidifiers are used with HWBS? 

 

Humidifiers used with HWBS are primarily classified based on their heating method and water chamber type:

 

  • Heated Humidifiers: These are the most common type, using a heating element to warm the water. They offer precise control over temperature and humidity. 
    • Pass-over Humidifiers: Gas flows over the surface of the heated water. 
    • Bubble Humidifiers: Gas is bubbled through the heated water (less common with ventilators, more common with lower-flow oxygen therapy). 
  • Water Chamber Types: 
    • Refillable Chambers: Require manual refilling with sterile water. 
    • Pre-filled, Disposable Chambers: Single-use chambers pre-filled with sterile water, reducing the risk of contamination. 
    • Auto-feed systems: The humidifier automatically refills from a larger sterile water source. 

 

 

How are Humidifiers and HWBS Classified and Grouped in Costa Rica? 

 

Both humidifiers and Heated Wire Breathing Systems (HWBS) fall under Class 2 medical devices according to Rule 5 of Decree 43902-S. 

 

Grouping Strategy for Registration: 

 

  • Separate Registrations: Humidifiers and HWBS require separate registrations with the Ministry of Health, even though they are often used together; they are distinct devices with different indications of use and classifications. 
  • Humidifier Grouping: Humidifiers with the same intended use, heating method, and similar materials can be grouped, even with variations in water chamber type (refillable vs. pre-filled), temperature control features. 
  • HWBS Grouping: Follows the guidelines outlined for regular Breathing Systems (similar materials, intended use, variations in length/connectors allowed). 

 

Specifications to be considered: 

 

    • Performance Testing: The humidifier and HWBS should undergo joint testing to confirm that the system consistently achieves the intended temperature and humidity levels without complications. 
    • Instructions for Use: Clear guidelines must be provided for both devices, detailing how to safely and effectively connect and operate them together. 
    • Sterility: Explicit definitions of sterility requirements and re-sterilization protocols must be established for each grouped device, including the HWBS kit.
    • Technical Documentation: The registration dossier must contain comprehensive technical specifications for all components of the system.
    • Components and Accessories: A rationale for including specific components and accessories is required, particularly when they are not packaged together or designed exclusively for use with the circuits. 

 

Conclusion 

 

Humidifiers and HWBS are essential components of modern respiratory care. Understanding their distinct classifications, the interaction between them, and the specific grouping strategies for registration in Costa Rica is crucial for manufacturers to ensure successful market access. 

 